Japanese   Deutsch   Portugues   Italiano   Francais      Espanol

OUR PRODUCTS

Microprocessor Soil Analysis Kits

We offer wide range of microprocessor soil analysis kits for measurement of pH, conductivity, mv, TDS, Salinity, D.O. & temperature 16 X 2 Alphanumeric LCD display. Battery Cum mains operated with all accessories fitted in a briefcase. Data storage facility up to 700 samples and external printer attachment facility available.

Microprocessor Soil Analysis Kit


Labtronics Products

Item Code: LT-59

We offer wide range of Microprocessor Soil Analysis Kit with following specifications.





Technical Specifications


pH:


Range              : 0-14 pH

Resolution         : 0.01pH

Accuracy           : + 0.01 pH ( relative)

Temp. Comp.        : Automatic with temp. Probe

Buffer : Manuals with keys

Recognition : 4.00 , 7.00 , 9.20 ( Automatic)

Slope Adjustment        : 80% to 120%, by default 100%

Calibration        : Auto/Manual 2-Point, 3-Point


Dissolved  Oxygen:

 

Range               : 0 to 20 ppm

Resolution          : 0. 1 ppm

Calibration         :  Automatic

Accuracy            : + 0.2 % + 1 digit  

Temp. Comp.         : Automatic with Pt 100 probe &manual with keys

Pressure Adjustment      : 0-999 mm of Hg ( Manual)

Salinity Correction : 0 to 20 gm/l (manual)


mv/ORP:

 

Range        : 0 to +1999 mV oC

Resolution   : 0.1 mV

Calibration  : Manual

Accuracy     : + 1mV


Temperature:

 

Range       : -20 to +200 oC

Resolution  : 0.1 oC

Accuracy    : +0.2%

Input Type      : Pt 100

Calibration :  Manual


Conductivity:

 

Range              : 0 to 1000 m mho in 6-ranges

Range Selection    : Automatic

Resolution         : 0.01 uS/cm

Accuracy           : + 0.5%

Temperature Comp.        : Automatic for entire range

Temp. Co-efficient : Adjustable between 0 to 4%

Cell Constant      :  Adjustable 


TDS:

 

Range             : 0 to 2000 ppt in 5 ranges

Range Selection   : Automatic

Resolution        : 0.1 ppm

TDS Factor          : 0.5 and adjustable manually


Salinity:

 

Range           : 0 to 50 ppt

Resolution      : 0.1 ppt

Temp. comp.     : Automatic

Accuracy        : +0.5% of range + 1 digit

Temperature Comp.        : Automatic

Salinity factor :  0.75 and adjustable manually

Display : 16  characters *2 lines alphanumeric LCD 

Display Format  :  Temp. and parameter  Simultaneously

Setting        :  By soft touch keys

Storage         :  1000 samples with Date & Time

Interface    :   RS-232, Centronics & Computer  

Real Time Display        : 24 hours mode with Date

Error Indication        : ph Error, Buffer Error, orp Error

Digital Output        : a)  Serial RS-232  b) Parallel Centronics

Print Format        : a)  All Data   b) Date Wise

Share on Facebook Share on Twitter Share on Myspace